VIDEO: Setting up the 3D SpectaSCOPE™ - Part 1

Adjustable Height/Working Distance

WPI's 3D SpectaSCOPE™ Inspection Scope is a true 3D digital stereo optical scope with a host of innovative patent-pending features.

It all starts with the fully optical zoom system, which is sending the smallest details to its two powerful HD 5MP cameras. The images are then transmitted at full speed frame rates to the on-board HD screens, giving you a true three-dimensional image. You choose the most comfortable position for the viewer’s tilting headset. 

The 3D SpectaSCOPE™ is easy to set up. Here we show you how to adjust the height and working distance of the 3D SpectaSCOPE™.

Step 1.   Use the adjustment knob to move the manual focus to the middle of the range.

Step 2.  While supporting the arm, loosen the tension adjustment knob/handle to move the head up or down.

Step 3.  Using a ruler, adjust the scope head so it is 200mm from the top of the subject you wish to view.

Step 4.  Tighten the tension adjustment knob/handle.

Step 5.  Looking through the visor, press the menu/zoom button.

Step 6. Rotate the zoom/menu button and select working distance from the menu options.

Step 7.  Select WD 200mm from the menu. The subject you wish to view should be nearly in focus.

Step 8. If necessary, Adjust the manual Focus (Step 1) to finalize focusing on the subject. The subject should now be in focus.


If you have any questions about the 3D SpectaSCOPE Inspection Scope, give us a call at (866) 606-1974 (Toll-free in the USA) or email us at [email protected].