MatTek PermaCell 96-well insert plates are advanced, sterile, cell culture devices used for growth and differentiation of cells. The 96-well insert plates consist of an array of 96 membrane wells connected into a single, rigid tray for easier handling and feeding
and to allow for high-throughput/ robotic processing. The 96-well insert plate fits within a 96-well receiver plate (included) which allows for individual treatment and assaying of each well in the plate.

Prices valid in USA, Canada, and PR only.

MatTek PermaCell 96-well insert plates can be used for cell culture and/or cell differentiation into 3-dimensional tissues, drug transport/permeability studies, and imaging studies.

NOTE: When working with MatTek PermaCell insert plates, use aseptic techniques and appropriate culture media, reagents, and growth environments for specific cell types. Depending on the specific insert and cell type, this could include coating the inserts with an extracellular matrix (ECM) or other attachment aids.

NOTE: WPI is an authorized distributor of MatTek for the 96-well cell culture insert plates.

Options

 Order code Description Membrane Type Pore Size Pore Density Transparent in Culture Insert Size Pkg.of
CCI96-PET-CL-0.4 PermaCell Culture Insert Plate, 96-well, Clear PET 0.4 μm 4x106 pores/cm2 Yes 96-well Pkg.of 1
CCI96-PET-CL-0.4-5 PermaCell Culture Insert Plate, Clear 96-well, Pack of 5 PET 0.4 μm 4x106 pores/cm2 Yes 96-well Pkg.of 5

DIRECTIONS FOR USE

To ensure sterility, all steps listed below should be performed in a laminar flow hood or equivalent controlled environment.
1. Inserts come pre-packaged in 96-well receiver plate.
2. Remove the insert plate and inspect the wells to ensure there are no defects.

COATING (OPTIONAL)

1. Coat insert plate with appropriate ECM following manufacturer’s recommendations
         a. 96-well plate inserts use 25-75 μL/insert

2. Incubate the plate following manufacturer's recommendations and then remove coating and wash inserts. The insert plates are now ready for use - refer to Plate Set-up below.

Plate Setup

1. Add cell culture media to wells of the 96-well receiver plate

            a. For MatTek 96-well insert plates, add 400 μL of medium into the wells of the receiver plate.

2. Seed cells into the wells of the insert plate and follow standard tissue culture incubation and feeding procedures for optimal cell growth.

             a. Typical experiments seed cells on the apical membrane using up to 150μL in the apical compartment.

             b. Experiments can also include seeding cells into the receiver plate wells and/or onto the basal side of the membrane.

NOTE: For TEER measurements, use the following media or buffer volumes.

TEER Measurement System Minimal Apical (top of the membrane) Volume, µL Minimal Basolateral (below the membrane or inside the well of the receiver plate) Volume, µL
Manual System (EVOMâ„¢ Manual with STX HTS) 150 400
Automated System (EVOMâ„¢ Auto) 150 550
More Information
SKU VAR-CCI96-PET-CL-0.4
Type Description
Dimensions 96-well Insert Plate
Reciever Plates 127.8 x 85.5 mm
Well Depth 12 mm
Membrane Diameter 4.29 mm
Membrane Area 0.14 cm2
Memrane Pore Size 0.4 μm
Solvent Compatibility Same as Polystyrene