Micro Nitric Oxide Sensors

As low as $1,122.00
Only %1 left
Prices valid in USA, Canada, and PR only.
Order code
Price range: $1,122 - $1,282
Prices valid in USA, Canada, and PR only.

The ISO-NOP007 has a tip diameter of just 7 microns and a length of 2mm. The response is linear over a wide dynamic concentration range of NO. The design is based on a single carbon fiber coated with WPI's NO-selective membrane. A detection limit of approximately 1 nM NO makes this electrode ideal for use in tissues and microvessels.

Prices valid in USA, Canada, and PR only.

7 and 30 µm sensors with exceptional performance – ideal for tissues and microvessels

   Don't forget the SNAP!
  • Excellent selectivity to NO
  • Rapid response time
  • Highly sensitive
  • For use with Apollo1000, Apollo4000, TBR4100 and TBR1025
  • Requires cable 91580 (sold separate)
  • Package of 3

Options

Order code Length Tip Diameter Shape
ISO-NOP007 2 mm 7 µm Straight
ISO-NOPF3005 0.5 mm 30 µm  Straight
ISO-NOPF3020 2 mm 30 µm  Straight

 Benefits

  • Linear response over a large dynamic concentration range of NO
  • Dry electrodes require no sleeves or filling solution

Applications

  • NO measurement in microvessels

The ISO-NOP007 has a tip diameter of just 7 microns and a length of 2mm. The response is linear over a wide dynamic concentration range of NO. The design is based on a single carbon fiber coated with WPI's NO-selective membrane. A detection limit of approximately 1 nM NO makes this electrode ideal for use in tissues and microvessels.

The ISO-NOP30 has a tip diameter of 30 microns and is available in two different tip lengths ISO-NOP3020 has tip length of 2 mm, ISO-NO3005 has tip length of 0.5 mm). The response is linear over a wide dynamic concentration range of NO. The design is based on a single carbon fiber coated with WPI's NO-selective membrane. A detection limit of approximately 1 nM NO makes these electrodes ideal for use in tissues and microvessels.

Nitric Oxide Sensor-Simultaneous measurement of force

The response of a 7μm NO sensor (ISO-NOP007) to successive additions of NO (100nM). Inset shows the linearity of the resulting calibration plot.

Nitric Oxide Sensor-changes of NO concentration

Simultaneous measurement of force (top trace) and changes of NO concentration (lower trace) in (A) the rat superior mesenteric artery relaxed with ACh and (B) a small human artery relaxed with ACh and SNAP. In this artery oxyhaemoglobin (oxyHb) partly reversed the increase in NO concentration with only a small change in force. [U. Simonsen, et al., J. Physiol., 1999, 516: 271-282.]

More Information
SKU VAR-2968
  ISO-NOP007 ISO-NOP3005 ISO-NOP3020
Outside Diameter  7 µm 30 μm 30 µm
Available Length 2 mm  0.5 mm 2 mm
Response Time < 3 seconds  < 3 seconds < 3 seconds
Lowest Detection Limit/Range 0.5 nM  1 nM 1 nM
Nominal Sensitivity-New sensor ≥1 pA/nM  ≥1.4 pA/nM ≥1.5 pA/nM
Baseline Drift none  none  none
Poise Voltage 865 mV  865 mV 865 mV 
Typical Quiescent Baseline Current, 25°C 300 pA  1000 pA 500 pA 
Acceptable Baseline Range 200-1500 pA  150-3500 pA 500-5000 pA 
Polarization Time 1+ hours  1+ hours 1+ hours 



Dantas, B. P. V, Ribeiro, T. P., Assis, V. L., Furtado, F. F., Assis, K. S., Alves, J. S., … Braga, V. A. (2014). Vasorelaxation induced by a new naphthoquinone-oxime is mediated by NO-sGC-cGMP pathway. Molecules (Basel, Switzerland), 19(7), 9773–9785. https://doi.org/10.3390/molecules19079773  