DNA/RNA Quantification Using DIPUV-Mini and a Tidas Spectrometer

AbstractDiptip mini

Concentrations of DNA in solution (31µg/mL and 688µg/mL) were measured with a spectrometer and UV/VIS light source in a DIPUV-Mini. Due to the 2mm pathlength, use of a DIPUV-Mini does not require a pre-measurement dilution within this concentration range, thus a potential source of error was eliminated.

Experimental Procedure

Close up of tip

Standard solutions of DNA (Sigma D1626) were prepared gravimetrically using 18.2MΩ/cm ultrapurified water as a solvent. Solutions were prepared between 0.0µg/mL and 687.6µg/mL.

Measurements were taken in triplicate using a DIPUV-Mini. The DIPUV-Mini was connected to a UV/VIS light source (WPI #D4H).

Data were collected in 1nm increments across the full range of the instrument (190nm-720nm). The instrument was configured such that reference measurements yielded an 80% total intensity. All measurements utilized 18.2MΩ/cm ultrapurified water as a reference solution.

 Results

Experimental results are presented in Table 1:

DNA [µg/mL] Absorbance @260nm  [AU]
0.00 -0.0017
30.83 0.1116
56.59 0.1945
85.49 0.3012
115.66 0.3821
143.94 0.4551
281.63 0.8399
428.91 1.2409
561.15 1.5639
687.61 1.7340

Table 1: DNA Concentrations and Resultant Absorbance Values

Since absorbance with respect to concentration follows the Beer-Lambert Law,

A= εlc
expected absorbance values were calculated from the DNA solution concentrations. Literature values of ε for dsDNA are listed as 0.020µg/ml*cm.

Experimental data can be found in Figure 1 with the calculated absorbance measurements indicated by a solid line. Absorbance measurements are expressed at a 260nm wavelength. Deviation from the theoretical value at higher absorbance values is a result of stray lightinterference within the spectrometer.

 Measured v. theoretical absorbance

 Figure 1: Measured versus Theoretical Absorbance

Typical DNA Measurement

Figure 2: Typical DNA Measurement (281.6µg/mL)