Stimulus Isolators / Precision

Current Sources

Activated by conventional logic-level commands, Model A360LA can be gated by any pulse generator, stimulator, or computer output.

Dual tone audible alarms - A tone sounds when an open electrode circuit is detected or when system compliance is reached. A second optional tone sounds when a signal is applied to the input. A test switch is also provided to check battery charge.

Current delivery - Stimulus currents are set using a three-digit control knob and a three-position range switch. Output current tracks control settings to better than 2% Output current is load independent; voltage sufficient to push the desired current through the load is automatically developed, subject only to compliance limits. Model A360LA produces up to 10 milliampere current, in three ranges, at more than 100 volts compliance.

Polarity - Output polarity is determined by a three-position switch on the front panel (+/-/off).

A360LA SPECIFICATIONS
Output Waveform DC or current pulse
Output Current Ranges 0.1, 1.0, and 10 mA
Current Amplitude Error 0.5% of full scale, max.
Current Resolution 0.1% of full scale, typical
Output Load Voltage Excursion (Compliance) 100 V minimum, 130 V nominal
External Command Threshold 2.2 V at 2.6 mA, min. 8.5 V, max. (TTL)
Output Polarity Reversible, manual switch
Current Rise Time & Delay 2 µs, typical (1 KΩ load) 5 µs @ 10 KΩ
Current Fall Time & Delay 5 µs, typical (1 KΩ load) 8 µs @ 10 KΩ
Output to Ground Resistance 1012 Ω
Optocoupler 2500 V, rated min. breakdown voltage
Noise <0.1 microampere (10 mA), < 50 nA (1 mA, 100 microampere A)
Current Leakage +/- 10 nA +/- 0.01% of full scale
Power Two 6-V lead/acid batteries, 2.4 Ah
Recharge Time 4 hours
Recharger Included 100-240 VAC, 50-60 Hz, universal
Dimensions 4.5 x 3.5 x 5.5 in. (11 x 9 x 14 cm)
Shipping Weight 4 lb (1.8 kg)
A365 Stimulus Isolator  

Model A365 includes the same features and specifications as A360 but with the added capability for bipolar pulsing for zero net charge.

Polarity - Output polarity is determined by a push switch on the front panel. Bipolar current is toggled by the command waveform, setting alternating pulses as positive or negative.

A365 SPECIFICATIONS
Output Waveform DC or current pulse
Output Current Ranges 0.1, 1.0, and 10 mA
Current Amplitude Error 0.5% of full scale, max.
Current Resolution 0.1% of full scale, typical
Output Load Voltage Excursion (Compliance) 100 V
External Command Threshold 2.2 V at 2.6 mA, min. 8.5 V, max.
Output Polarity Reversible, manual switch or automatic
Current Rise Time & Delay 6 µs, typical (1 KΩ load)
Current Fall Time & Delay 10 µs, typical (1 KΩ load)
Output to Ground Resistance 1012 Ω
Optocoupler 2500 V, rated min. breakdown voltage
Power Model A365D (dry cell) 16 alkaline 9 V batteries, included
Model A365R (rechargeable) 16 rechargeable NiMH 9 V batteries incl.
Dimensions 8.5 x 3.5 x 5 in (22 x 9 x 12 cm)
Shipping Weight 4 lb (1.8 kg)
Order Numbers
A365RC A365R with Charger (A362)

A360-LA

High Voltage Isolator, lead acid battery, charger included

Specify line voltage

A365D

High Voltage Isolator, Bipolar, standard battery

A365R

High Voltage Isolator, Bipolar, rechargeable

A362

Battery Charger for A320R, A360R,

Specify line voltage

OPTIONAL ACCESSORIES

DRL

Dummy Load Resistor Kit (set of 3)

3468

Dual Rack Mount Kit for A365

3469

Single Rack Mount Kit for A365