TIDAS I Specifications
|
| Optics |
| Optical Basics |
Monolithic spectrometer module, Single Beam System. Fiber optic cross section for increased light throughput instead of entrance slit. Zeiss multilayer filter. Hamamatsu photodiode array detector, 16 Bit
|
| Wavelength range |
195 730 nm
|
| Number of diodes |
256
|
| Resolution |
2.2 nm/pixel
|
| Spectral bandwidth |
(FWHM) 5 nm
|
| Wavelength accuracy |
+/- 0,2 nm
|
| Wavelength reproducibility |
+/- 0,07 nm
|
| Noise (at 550 nm) |
+/- 0.5 x 10-5 AU (using FO-6000)
|
| Dimension (h x w x d) |
6.3 in. x 10.6 in. x 12.8 in.
160 mm x 270 mm x 325 mm)
|
| Electrical requirements |
90-240V, 50/60Hz
|
| Weight |
18 lb (8 kg)
|
| Electronics and Data processing
|
| PC Interface |
Serial RS 232 Interface
|
| System |
350 Mhz PIII, 64Mbyte, Windows 98/2000, Windows NT.
|
| Digital inputs/outputs |
5/5
|
| Spectralys Software (for Windows) |