TIDAS I Specifications
Optics
Optical Basics Monolithic spectrometer module, Single Beam System. Fiber optic cross section for increased light throughput instead of entrance slit. Zeiss multilayer filter. Hamamatsu photodiode array detector, 16 Bit
Wavelength range 195 – 730 nm
Number of diodes 256
Resolution 2.2 nm/pixel
Spectral bandwidth (FWHM) 5 nm
Wavelength accuracy +/- 0,2 nm
Wavelength reproducibility +/- 0,07 nm
Noise (at 550 nm) +/- 0.5 x 10-5 AU (using FO-6000)
Dimension (h x w x d) 6.3 in. x 10.6 in. x 12.8 in. 160 mm x 270 mm x 325 mm)
Electrical requirements 90-240V, 50/60Hz
Weight 18 lb (8 kg)
Electronics and Data processing
PC Interface Serial RS 232 Interface
System 350 Mhz PIII, 64Mbyte, Windows 98/2000, Windows NT.
Digital inputs/outputs 5/5
Spectralys Software (for Windows) Absorption, Transmission, Counts, Concentration, Multi-Calibration, Multiwavelength measurement. Spectral smoothing, Derivative (up to D 4), Macro-programming.