TIDAS II Specifications |
| Optical Basics |
Diode-Array-Spectrometer with ceramic Chassis. Single Beam System. Fiber optics beam shaper instead of entrance slit. Zeiss multilayer filter. Optimized for fiber optics. External fiber optics coupling free off loss of light switched by controlling software
|
| Light sources |
35 W deuterium - 7.5W tungsten lamp. Light mixed with Y-coupler.
|
| Wavelength range |
187-1023 nm
|
| Measuring speed |
12 ms/Spectrum
|
| Number of diodes |
1024
|
| Resolution |
0.8 nm/Pixel
|
| Spectral bandwidth |
2,0 nm
|
| Wavelength accuracy |
+/- 0,1 nm
|
| Wavelength reproducibility |
+/- 0,07 nm
|
| Noise at 550 nm |
+/- 2,5 x 10-5 AU
|
| Baseline drift |
5 x 10-4 AU/h
|
| PC-Interface |
RS 422 Interface (included) |
| System Requirements |
350 MHz PIII, 64 Mb, Windows 98/2000 |
| Digital inputs/outputs |
1/1 |
| Dimensions |
18.5 x 49 x 47 cm
|
| Electrical requirements |
120 / 240 V, 60 / 50 Hz
|
| Weight |
41 lbs (18.5 kg) |