TIDAS II Specifications

Optical Basics Diode-Array-Spectrometer with ceramic Chassis. Single Beam System. Fiber optics beam shaper instead of entrance slit. Zeiss multilayer filter. Optimized for fiber optics. External fiber optics coupling free off loss of light switched by controlling software
Light sources 35 W deuterium - 7.5W tungsten lamp. Light mixed with Y-coupler.
Wavelength range 187-1023 nm
Measuring speed 12 ms/Spectrum
Number of diodes 1024
Resolution 0.8 nm/Pixel
Spectral bandwidth 2,0 nm
Wavelength accuracy +/- 0,1 nm
Wavelength reproducibility +/- 0,07 nm
Noise at 550 nm +/- 2,5 x 10-5 AU
Baseline drift 5 x 10-4 AU/h
PC-Interface RS 422 Interface (included)
System Requirements 350 MHz PIII, 64 Mb, Windows 98/2000
Digital inputs/outputs 1/1
Dimensions 18.5 x 49 x 47 cm
Electrical requirements 120 / 240 V, 60 / 50 Hz
Weight 41 lbs (18.5 kg)