We have upgraded these products. For assistance please call 1-941-371-1003
CHOOSING A GRATING
SPECIFICATIONS
ORDERING INFORMATION CHOOSING A SLIT
WPI's Miniature Fiber Optic Spectrometers are high-performance, low-cost instrument designed for use in the UV/VIS range. The SMA fiber optic design permits easy configuration into many applications to measure absorbance, reflectance, radiance, fluorescence and other optical properties. The system is capable of high precision light measurements over more than six orders of magnitude. The S2000 spectrometers are available with 3 popular gratings.
S2000
FIBER OPTIC SPECTROMETER SPECIFICATIONS
DETECTOR 2048-element CCD linear array
CCD Elements 2048 x 12.5 µm x 200 µm
Well Depth 350,000 photons
SENSITIVITY 86 photons/count
2.9 x 10-17 joule/count
2.9 x10-17 watts/count (for 1-sec integration)
Effective Range 200 - 1100 nm
Integration Speed Å 7.2 - 8 milliseconds to 4 seconds (with 500 kHz A/D card)
OPTICS
Gratings 14 grating choices, optimized for UV, VIS, or NIR
Slits 10, 25, 50, 100, or 200 µm x 1000 µm
Order-Sorting Filters Permanently installed Schott glass longpass filters
Resolution Å 0.3 - 10.0 nm FWHM (depends on groove density of grating and diameter of fiber or width of slit)
Stray Light < 0.01% at 600 nm
< 0.10% at 435 nm
Fiber Optic Connector SMA905 connector to single strand optical fiber (0.22 NA)
Order Numbers
S2000SGL Single Beam, SpectraWare, A/D card (PCI), cables
S2000DBL Dual Beam / SpectraWare, A/D card (PCI), cables

L2

Low Light Upgrade: Detector Collection Lens for S2000, installed per channel
UV2 UV Detector Upgrade for S2000 Series, required for measuring at wavelengths less than 350 nm, installed per channel
OFLV-200-850 Multi-bandpass order-sorting detector coating, installed per channel
OFLV-350-1000 Multi-bandpass order-sorting detector coating, installed per channel
SLIT-xxx (other than 50 micron) 10, 25, 100, 200 /per channel