Eagle Pipetter Specifications
|
|
Model |
Volume Range |
Dispensed Volume
|
Accuracy |
Repeatability |
|
|
0.2 µL |
± 12 % |
( ± 0.024 µL) |
± 6 % |
( ± 0.012 µL) |
|
E2 |
0.2-2 µL |
0.5 µL |
± 2.7 % |
( ± 0.0135 µL) |
± 1.3 % |
( ± 0.0065 µL) |
|
|
2 µL |
± 1.5 % |
( ± 0.03 µL) |
± 0.7 % |
( ± 0.014 µL) |
|
. |
. |
1 µL |
± 2.5 % |
( ± 0.025 µL) |
± 1.3 % |
( ± 0.013 µL) |
|
E10 |
0.5-10 µL |
5 µL |
± 1.5 % |
( ± 0.075 µL) |
± 0.6 % |
( ± 0.03 µL) |
|
|
10 µL |
± 1 % |
( ± 0.1 µL) |
± 0.4 % |
( ± 0.04 µL) |
|
|
|
2 µL |
± 7.5 % |
( ± 0.15 µL) |
± 2 % |
( ± 0.04 µL) |
|
E20 |
2-20 µL |
5 µL |
± 5 % |
( ± 0.25 µL) |
± 1.5 % |
( ± 0.075 µL) |
|
|
10 µL |
± 3 % |
( ± 0.3 µL) |
± 1 % |
( ± 0.1 µL) |
|
|
20 µL |
± 3 % |
( ± 0.6 µL) |
± 0.8 % |
( ± 0.16 µL) |
|
|
|
20 µL |
± 3 % |
( ± 0.6 µL) |
± 1 % |
( ± 0.2 µL) |
|
E100 |
20-100 µL |
50 µL |
± 2 % |
( ± 1 µL) |
± 0.5 % |
( ± 0.25 µL) |
|
|
100 µL |
± 1.5 % |
( ± 1.5 µL) |
± 0.3 % |
( ± 0.3 µL) |
|
|
|
50 µL |
± 3 % |
( ± 1.5 µL) |
± 1 % |
( ± 0.5 µL) |
|
E200 |
50-200 µL |
100 µL |
± 1.3 % |
( ± 1.3 µL) |
± 0.7 % |
( ± 0.7 µL) |
|
|
200 µL |
± 1.3 % |
( ± 2.6 µL) |
± 0.5 % |
( ± 1 µL) |
|
|
|
100 µL |
± 2 % |
( ± 2 µL) |
± 0.8 % |
( ± 0.8 µL) |
|
E1000 |
100-1000 µL |
500 µL |
± 1 % |
( ± 5 µL) |
± 0.4 % |
( ± 2 µL) |
|
|
1000 µL |
± 0.8 % |
( ± 8 µL) |
± 0.28 % |
( ± 2.8 µL) |
|
|
E5000 |
1000-5000 µL |
1000 µL |
± 1.3 % |
( ± 13 µL) |
± 0.4 % |
( ± 4 µL) |
|
|
2000 µL |
± 0.8 % |
( ± 16 µL) |
± 0.3 % |
( ± 6 µL) |
|
|
5000 µL |
± 0.8 % |
( ± 40 µL) |
± 0.2 % |
( ± 10 µL) |
|