Pneumatic PicoPumps
Pressure-pulsed micro injector for volumes ranging from picoliters to milliliters
|
Excellent for focal drug delivery
Simplify intracellular injection
Regulated air pressures for securing cells and injecting with fluid
Injected volumes range from picoliters to milliliters
Positive pressure for high-pressure ejection, suction for supporting the cell or filling the pipette from the tip
Negative pressure port maintains a low positive holding pressure to the injecting pipette between injection pulses preventing fluid uptake through capillary action or diffusion.
Adjust timing, ejection pressure, holding pressure, and suction independently
Injection pressure is controlled by a 20 turn regulator on the front panel.
Built-in timing circuit allows precise control of the amount of time that the injection pressure is applied to the output port. Time intervals can range from 10 seconds down to 10 ms or less, depending on the eject pressure setting.
The injection pressure interval can be triggered manually on the front panel, by footswitch, or by computer controlled TTL pulse.
A 5-volt monitor output provides a logic-level pulse for your computer or other monitoring device.
|
 |
 |
PV830
Eject Pressure, Hold Pressure, and Vacuum are all available, controlled by separate regulators on the front panel.
Eject Pressure supplies a high-pressure pulse for injecting fluid.
Hold Pressure, which is not sufficient to cause fluid ejection, is used to prevent back filling of the pipette by capillary action or diffusion when the solenoid is inactive. Pressure in the injection pipette is automatically switched between Eject and Hold Pressure by a precision timing circuit that controls a solenoid valve.
Eject Pressure is controlled by the range switch and 20-turn regulator knob on the front panel. Vacuum is used to fill pipettes from the tip or to secure a floating cell during microinjection.
Vacuum is regulated the same way, by a 20-turn knob on the front panel. Vacuum may be switched from regulated vacuum to atmosphere by using the pneumatic toggle on the front panel. Ejection or suction can be timed manually, using the Start button or a footswitch (optional), or timing can be preset on the front panel by using the Range switch and Period knob, or by computer.
|
PV820
Eject Pressure and Hold Pressure are available on this unit.
Hold Pressure is
used to maintain a low pressure in the pipette between injections to prevent unwanted fluid uptake by capillary action or diffusion.
A precision timing circuit switches from Eject Pressure to Hold Pressure automatically, once timing has been set.
Add vacuum externally
Although regulated vacuum is not provided in this model, suction can be provided by connecting a vacuum source to the vacuum port on the rear panel. Suction is then available through the pressure ports. Ejection/suction can be timed manually, using the Start button or a footswitch (optional).
Timing can also be preset by using the Range switch and Period knob on the front panel, or by computer.
|
PicoNozzle Kit (5430-ALL) included with both the PV820 & PV830 allows all sizes of glass micropipettes to be securely mounted in micropositioners for stable axial air delivery. Because air enters the pipette axially, lateral whipping during injection is eliminated.
 |
PicoNozzle Kit 5430-XX (see below) - Available as an accessory, choose the kit specifically designed for the size of glass capillaries used in your application.

|
|
References PV820 Morishima, M, et al. "Recurrent Connection Patterns of Corticostriatal Pyramidal Cells in Frontal Cortex" (2006) The Journal of Neuroscience 26(16)4394-4405 Antonsen, B, et al. "The Retrograde Spread of Synaptic Potentials and Recruitment of Presynaptic Inputs" (2005) The Journal of Neuroscience 25(12):3086-3094
He, D, et al. "Prestin and the Dynamic Stiffness of Cochlear Outer Hair Cells" (2003) The Journal of Neuroscience 23(27):9089-9096
PV830
Griffioen, K, et al. "Purinergic P2X Receptors Mediate Excitatory Transmission to Cardiac Vagal Neurons in the Nucleus Ambiguus After Hypoxia" (2007) Hypertension 50:75-81
Wang, X, et al. "Ketamine Inhibits Inspiratory-evoked y-Aminobutyric Acid and Glycine Neurotransmission to Cardiac Vagal Neurons in the Nucleus Ambiguus" (2005) Anesthesiology 103:353-9
Deveau, J, et al. "An improved method for contructing and selectively silanizing double-barreled, neutral liquid-carrier, ion-selective electrodes" (2005) Bio Proced. Online 7(1):31-40
|
|
Using PV830 for holding and injection of cells
view application sample
For a complete list of pre-pulled micropipettes, see µTips,
or contact us with your special requirements.
Request a quote |
Saitow, F, et al. "Beta-Adrenoceptor-mediated long-term up-regulation of the release machinery at rat cerebellar GABAergic synapses" (2005) J Physiol 565.2:487-502
Wang, J, et al. "Developmental Effects of Ectopic Expression of the Glucocorticoid Receptor DNA Binding Domain Are Alleviated by an Amino Acid Substitution That Interferes with Homeodomain Binding" (1999) Molecular and Cellular Biology 7106-7122
Gutierrez, A, et al. "Activation of a Ca2+-permeable cation channel by two different inducers of apoptosis in a human prostatic cancer cell line" (1999) J. Physiol. 517:95-107
Gomez-Viquez, L, et al. "SERCA Pump Optimizes Ca2+ Release by a Mechanism Independent of Store Filling in Smooth Muscle Cells" (2003) Biophysical Journal 85:370-380 |
| PV820 |
PicoPump w/ hold pressure |
| PV830 |
PicoPump w/ hold pressure and vacuum |
| Specify line voltage. All PicoPumps require external vacuum source -- see below |
| Each PicoPump is supplied with one PicoNozzle 2 Kit (5430-ALL - package of 2), for glass ODs of 1.0, 1.2, 1.5, or 1.65 mm, plus tubing to connect the holders to the pressure and vacuum ports as applicable |
|
| |
|
| Optional Acessories |
|
3260 |
Foot Switch |
| 2932 |
Rack Mount Kit, 31/2-in. high (PV820) |
| 2933 |
Rack Mount Kit, 51/4-in. high (PV830) |
| 5430-ALL |
PicoNozzle (V.2) Kit (for glass ODs of 1.0, 1.2, 1.5, or 1.65 mm) plus tubing |
| 5430-10 |
PicoNozzle Kit (MPH6S for 1.0 mm pipette & 5-ft tubing assembly) |
| 5430-12 |
PicoNozzle Kit (MPH6S for 1.2 mm pipette & 5-ft tubing assembly) |
| 5430-15 |
PicoNozzle Kit (MPH6S for 1.5 mm pipette & 5-ft tubing assembly) |
| 5430-20 |
PicoNozzle Kit (MPH6S for 2.0 mm pipette & 5-ft tubing assembly) |
| MPH6S |
Micropipette Holder (specify 1.0, 1.2, 1.5 or 2.0 mm) |
| MPH6R |
Micropipette Holder (specify 1.0, 1.2, 1.5 or 2.0 mm) |
| 3316 |
Replacement Input Kit |
| LV140-Y |
Vacuum Pump (120V only) |
|
PV800 SERIES SPECIFICATIONS
|
|
PV820 |
PV830 |
| PRESSURE |
| Pressure Input |
0 to 150 psi |
0 to 150 psi |
| Pressure Output |
0.3 to 90 psi * |
0.3 to 90 psi |
| Lowest Regulated Pressure |
12 in. water * |
12 in. water |
| Regulator Accuracy |
0.1% (20-turn dial) * |
0.1% (20-turn dial) * |
| Regulator Repeatability |
0.05 psi * |
0.05 psi * |
| Gauge Accuracy |
3% at full scale * |
3% at full scale * |
| Input Connector |
Quick Connect (1/4 in. OD Tubing) |
Quick Connect (1/4 in. OD Tubing) |
| Output Connector |
Barbed (1/16 in. ID Tubing) |
Barbed (1/16 in. ID Tubing) |
| Control |
Solenoid |
Solenoid |
| Vent |
To holding pressure or vacuumby manual control
* Both Hold and Eject Pressures * Both Hold and Eject Pressures |
To vacuum, holding pressure or atmosphere |
| VACUUM |
| Vacuum Input |
0 to 30.0 in. Hg |
0 to 30.0 in. Hg |
| Vacuum Output |
Unregulated |
0.2 to 29.9 in. Hg |
| Lowest Regulated Vacuum |
Unregulated |
3 in. water |
| Regulator Accuracy |
Unregulated |
0.1% (20-turn dial) |
| Regulator Repeatability |
Unregulated |
0.03 in. Hg |
| Gauge Accuracy |
None |
3% at full scale |
| Input Connector |
Quick Connect (1/4 in. OD Tubing) and Barbed (3/16 in. ID Tubing) |
Quick Connect (1/4 in. OD Tubing) and Barbed (3/16 in. ID Tubing) |
| Output Connector |
Barbed (1/16 in. ID Tubing) |
Barbed (1/16 in. ID Tubing) |
| Control |
Manual |
Manual |
| Vent |
Atmosphere |
Atmosphere |
| CONNECTIONS INCLUDED |
| Input Kit |
12 ft nylon tubing (0.25 in. OD, 1000 psi), two 1/4-in. OD adapter, one female NPT adapter |
| Output Kit |
Two PicoNozzle assemblies, each consisting of one MPH6S pipette holder, 60-in. of PVC tubing (200 psi), and a luer-fitted aluminum handle. |
| TIMING |
| Duration Modes |
Gated (by input signal) or Timed (by internal clock) |
| Pulse Initiation |
Manual, External Input or Foot Switch (optional) |
| Pulse Width |
10 ms to 10 s in Timed Mode (10-turn dial). Lower limit depends upon pressure setting. At 0 psi the minimum pulse width is nominally 8 ms. At 100 psi the minimum pulse width is nominally 3 ms. |
| Accuracy |
0.1% at full scale |
| External Input |
+5 V TTL-compatible (BNC) |
| Monitor Output |
+5 V TTL-compatible (BNC) |
| PHYSICAL SPECIFICATIONS |
| Power |
95-135 V or 220-240 V, 50/60 Hz |
95-135 V or 220-240 V, 50/60 Hz |
| Dimensions |
17 x 3.5 x 9.5 in. (43 x 9 x 24 cm) |
17 x 5.25 x 9.5 in. (43 x 13 x 24 cm) |
| Shipping Weight |
11 lb (5 kg) |
14 lb (6.3 kg) |
|
|
|
|