Microprobe Thermometer Specifications
|
| |
BAT 10 |
BAT 12 |
| Temperature Range & Resolution |
-200°C to +400ºC, 1ºC resolution |
-100°C to +199.9ºC, 0.1ºC resolution |
| |
-100°C to +199.9ºC, 0.1ºC resolution |
|
| Differential Temp. Range |
-19.99°C to +19.99ºC |
N/A
|
| |
Linearization centered at 40°C |
|
| |
0.01°C resolution
|
|
| Accuracy |
|
|
|
1° Range
|
1°C ±1 least significant digit |
0.1°C ± 1 digit between 0-50°C |
|
0.1° Range
|
0.1°C ± 1 least significant digit |
0.1% ± 1 digit over full range |
|
Diff. Range
|
0.01°C ± 1 least significant digit |
|
|
Repeatability
|
± 1 least significant digit
|
|
| Calibration Conformity |
Conforms to NIST tables |
Follows NIST thermocouple tables within 1 digit |
| Display |
31&Mac218;2 Digit LCD |
31&Mac218;2 Digit LCD |
| Input Socket |
Miniature, quick disconnect, copper-constantan |
Miniature, quick disconnect, copper-constantan |
Analog output
|
Non-linearized set at 1.6 V, corresponding to temperature of 401°C |
approx 10 mV per degree C
|
Power Supply/Batteries
|
BAT-10: 4 alkaline C cells (life: 1000 hr)
BAT-10R: 4 Ni-Cad C cells (rechargeable unit)
|
BAT-12: 9V cell
BAT-12R: 9V Ni-Cad with charger
|
Sensors
|
One WPI Type T thermocouple input
Ambient Operating Range: 15-45°C
|
Three WPI Type T thermocouple inputs |
|
|
|
Ambient Temperature Compensation:
Auto-compensated to 0.1°C from 0°C to 50°C
|
| Dimensions |
21.6 x 22.9 x 8.9 cm (8.5 x 9 x 3.5 in.) |
12.7 x 6.4 x 15.2 cm (5 x 2.5 x 6 in.) |
| Weight |
1.6 kg (3.5 lb), including carrying case |
1 kg (2 lb), including carrying case |