DAM series Bioamplifier

A family of very low battery operated amplifiers

WPI's DAM series amplifier’s are well known as a standard of the industry for extracellular potential amplification. These battery powered bio-amplifiers are designed with a compact chassis profile that enables the user to locate the unit closer to the preparation and thereby minimize long lead lengths which contribute to noise. Each amplifier is equipped with selectable high and low filters, and a position control to offset galvanic potentials which may develop during recording. A choice of models offer additional features that are useful for certain applications:

DAM50 - Basic Amplifier (optional #5447 electrode adapter not include) DAM80 - With low-noise headstage DAM80P (Metal microelectrode adapter #5469 included.)
Application Examples
DAM80, an AC amplifier only, features a very low noise headstage probe which can be mounted in micromanipulators for up-close cortical recording, for extracellular recording from high impedance glass or metal microelectrodes. Also provides a gated cur­rent for tissue marking. Microelectrode holder MEH3SB is recommended.
Feature DAM50 DAM80
Input mode AC/DC AC
Input configuration Differential/single ended Differential
Gain Range 100-10K (AC), 10-1K (DC) 100-10K (AC)
High/Low filters Y Y
Offset position control Y Y
Current Generator N Y
Stimulus Artifact Suppression N N
Remote Active headstage N Y
Output Connection BNC 3.5 mm mini phone
Standard input connection* unterminated wire mini banana
Power Supply (2) nine volt alkaline batteries (2) nine volt alkaline batteries
*See optional accessories for additional alternatives
DAM SERIES SPECIFICATIONS
INPUT IMPEDANCE 1012 Ohms, common mode and differential
INPUT LEAKAGE CURRENT 50 pA (typical)
MAX. DC DIFFERENTIAL SIGNAL ± 2.5 V (DAM 50)
GAIN AC: 100x, 1000x, 10000x
DC: 10x, 100x, 1000x (DAM50)
COMMON MODE REJECTION RATIO 100 dB @ 50/60 Hz
INPUT APACITANCE 20pF
 AC Mode Noise 0.4 µV RMS (2 µV p-p) 0.1-100 Hz
 AC Mode Noise 2.6 µV RMS (10 µV p-p) 1 Hz-10 kH
 DC Mode (DAM50) 7.5 µV RMS (30 µV p-p) 3-10 kHz
BANDWIDTH FILTER SETTINGS
 AC Mode Low frequency, 0.1, 1, 10, 300 Hz
 AC Mode (DAM80) High frequency, 0.1, 1, 3, 10 kHz
 DC Mode (DAM50) High frequency, 0.1, 1, 3, 10 kHz
OUTPUT CONNECTIONS BNC on DAM50; 3.5 mm Miniphone connection on DAM80
OUTPUT VOLTAGE SWING ±8 V
OUTPUT RESISTANCE 470 Ohms
BATTERY TEST Audible tone
CALIBRATOR SIGNAL 10 Hz square wave
POSITION Approx. 250 mV
CURRENT SOURCE
 DAM80: DC Generator 0 to ±50 µA, variable
EXTERNAL COMMAND Input Voltage ±10 V commands
 AC or DC current waveform ±50 µA max. amplitude @ 200 K Ω
BATTERIES 2 x 9 V alkaline (included)
DIMENSIONS
 DAM50 8 x 4 x 1.75 in. (20.3 x 10.2 x 4.4 cm)
 DAM80 7 x 4 x 1.75 in. (17.8 x 10.2 x 4.4 cm)
SHIPPING WEIGHT 3.5 lb (1.6 kg)
DAM50 Bio-amplifier
DAM80 Bio-amplifier with active probe (DAM80P)
OPTIONAL ACCESSORIES
DAM80P Replacement Probe
3072 6 Replacement Modular Cables
3517 2 Optional Shielded Modular Cables
CBL102 3.5 mm Phone plug-to-BNC Cable
2033 Black Insulated Mini-Banana Plug
2034 Red Insulated Mini-Banana Plug
2035 Uninsulated Mini-Banana Plug
2101 9V Alkaline Battery, each (2 required)
3484 Rack Mount Kit (for 2 DAM preamps)
3485 Ringstand Mounting Kit
5447 Electrode Adapter
5469 Metal Microelectrode Adapter for DAM80 (mini-banana plug to 0.031 in. (0.79 mm) socket)
5489 Adapter for Metal Microelectrode (DAM50)
3578 Adapter Cable for Ag/AgCI pellets (2 mm pin)
300102 Electrode Extention 4-inch
3414 9V NiMH Battery
13388 Adapter, mini-banana plug to 2mm socket