Features
Activated by conventional logic-level commands, Model A365 can be gated by any pulse generator, stimulator, or computer output.
Dual Tone Audible Alarms
A tone sounds when an open electrode circuit is detected or when system compliance is reached. A second optional tone sounds when a signal is applied to the input. A test switch is also provided to check battery charge.
Current Delivery
Stimulus currents are set using a three-digit control knob and a three-position range switch. Output current tracks control settings to better than 1%. Output current is load independent; voltage sufficient to push the desired current through the load is automatically developed, subject only to compliance limits. Model A365 produces up to 10 milliampere current, in three ranges, at more than 100 volts compliance.
Polarity
Output polarity is determined by a three-position switch on the front panel (+/-/off).
Bipolar Pulsing
Bipolar current is toggled by the command waveform, setting alternating pulses as positive or negative.
To place an order for this product, please contact the appropriate WPI office or distributor.
Customers from North America can order direct from our corporate office:
World Precision Instruments
175 Sarasota Center Boulevard
Sarasota, FL 34240-9258
- Phone or Fax
Toll-free: 866-606-1974 Tel: 941-371-1003 Fax: 941-377-5428
Our Customer Service Representatives are available 8am - 6pm EST.
- Online Store
Many of our products are available for purchase through our online store.
At this time we can only accept orders from the U.S. through the online store.
-
Want instant pricing? Order a free catalog here
International customers, please see our
office and distributor contact information list.