Isolated Stimulation and Stimulus Isolators
The term stimulation refers to the delivery of energy of some kind to a
biological tissue in order to elicit an observable response.
Although the energy used in stimulation may be chemical, thermal,
mechanical or electrical, this discussion will focus on electrical
stimulation. Electrical stimulation of biological tissues involves the
delivery of current and voltage to the stimulation site. The two
quantities are related by Ohm's law:
V=IR
Where V is the applied voltage, I is the current and R is the electrical resistance
of the tissue and or the stimulating electrodes. This simple equation
shows that if voltage is constant, current flow will diminish if the
tissue/electrode resistance goes up and will increase if the resistance
decreases.
More commonly, the resistance of tissue differs from sample to sample,
and the resistance of the electrodes changes with applied current over
time in a process called polarization.
Types of Stimulus Devices
In the delivery of electrical energy to biological tissues, stimulus
devices can hold either current or voltage constant during the
stimulating process. Devices that hold voltage constant at a value set
by the user and allow current to be determined by Ohm's law are known as
constant voltage stimulators.
Devices that hold current constant at a value set by the user during the
stimulation process and allow voltage to be determined by Ohm's law are
called constant current stimulators.
Constant current stimulators are preferred for two reasons:
- First, current is the quantity that stimulates most excitable tissues.
- Second, stimulating electrodes tend to increase their resistance as
stimulation progresses, as do some tissues. A constant current
stimulator will "sense" resistance change and provide whatever voltage
is needed to maintain delivery of current at the set rate.
There is obviously a limit to how much voltage a constant current
stimulator can provide. If the resistance of the preparation becomes infinite, as might happen if one of the stimulating electrodes is removed from the tissue, the stimulator could not mount an infinite
voltage to compensate.
The maximum amount of voltage that a constant current stimulator can
provide is called the compliance voltage. Once this compliance voltage
has been reached, further increases in tissue resistance will cause a
drop in delivered current.
WPI isolators in the 300 and DLS series offer a compliance voltage of 100V with very low noise. Stimulus isolators, as the name implies, also isolate a given stimulus
from ground. In an instrument design context most people think of
isolation from ground as it relates to electrical safety. From a
biological recording standpoint there are other issues. Consider the
circuit in Figure 1.
The stimulator in this case is a battery with a switch; current leaves
the positive terminal of the battery, travels down the stimulating lead,
passes through the tissue, and 100% of the current delivered returns to
the negative terminal of the battery. This figure also shows a
voltmeter in the preparation.
Voltages generated by the tissue as a consequence of the stimulus are recorded with respect to the voltmeter's ground electrode.
In Figure 2 the battery is replaced with a line powered stimulator. Even
though the stimulus source and the voltmeter have separate ground
electrodes, they represent the same electrical point. For this reason a
significant portion of the stimulus current returns to ground by way of
the voltmeter's ground lead.
If the currents are significant, or the voltages that you are trying to
measure are very small, the I x R drop across the resistance of the
voltmeter's ground electrode will add to the recorded voltage from the
tissue and will be seen as a DC artifact.
Capacitative coupling between the voltmeter circuit and the isolated
circuit can induce current to flow in the voltmeter ground. The induced
current will transiently flow across the resistance of the voltmeter
ground and its I x R drop will be seen as a transient spike in front of
and behind any pulse of current delivered by the current source. This is
termed an AC artifact.
The primary reason why researchers use an isolated current source is to
minimize artifact. But, what if you are not recording? There is no
voltmeter ground to produce an observable artifact. If you stop and
consider the examples above, the artifact was minimized or eliminated by
controlling the path of the stimulus current.
Isolated Stimulation / Stimulus Isolators
Knowing the current path can be critical in physiological stimulation.
Consider the current path in Figure 3. The animal represented by the
badly drawn cat is secured in a stereotaxic frame.
The frame is grounded. The animal contacts the frame at multiple points.
Figure 3 shows the battery and switch model of a stimulator. Here, as
before, all of the current that leaves the positive side of the battery
must return to the negative side. 100% of the stimulus current must pass
between the stimulus electrodes. In this case the path as well as the
exact amount of current delivered is known.
Figure 4 shows the same experimental setup except the battery has been
replaced by a line powered voltage source. Stimulus current now returns
to ground via many routes. The amount of current that flows to ground is
determined by the resistance between the source and each of the ground
points and is calculated as a resistive network using Kirtchoff's laws.
Many unintended areas of the animal may be stimulated.
In the real world we cannot use a battery and a switch, particularly if
the current durations are on the order of milliseconds. Electronic
devices such as pulse generators and computers are used to generate
timing, and Isolators driven by these devices are used to deliver the
stimulus.
By connecting an isolator (even a battery powered one) to a mains
powered pulse generator, you connect the isolated ground of the isolator
to the mains ground of the pulse generator. Unless the electrical
connection between the two devices is accomplished without using a
mechanical connection between the two devices you break the isolation.
This is what makes an isolator an isolator. The non-mechanical contact
between machines that constitutes the isolation barrier can be
accomplished in one of two ways.
Originally, isolators were transformer isolated. Pulse waves were
applied to the primary winding of a transformer, while the actual
stimulus was derived from the secondary winding. The transference was
accomplished by induction. This approach suffered from two shortcomings.
The devices could not pass DC, so no constant voltage isolated
stimulations could be made. The transformer approach also has
intrinsically higher capacitance. This means that while the resistance
between the primary and secondary coils is very high, the high
capacitance produces an AC artifact that is unacceptably large
compared to other isolation techniques.
Optical isolation is the second popular scheme. The field has all but
standardized on this. In simple terms, the input pulse wave powers a
light that shines across the barrier onto a photocell that produces the
stimulus wave. There have been countless variations on this theme, and it
is used for isolating recording amplifiers, as well as stimulators.
WPI's Isolators (All are Optical)
DLS Series
WPI's digital stimulator, DS8000 has advanced features and specifications that are not found in any other digital stimulator.
The DLS100 is a revolutionary new breed of digital linear isolators with high compliance and high isolation. The DLS100 is preferentially optimized for applications with the DS8000 via a flexible cable through which it receives power and stimulus signals in a digital format. Up to eight DLS100 isolators can be connected independently to one DS8000. Very high isolation is achieved through the use of optical coupling of
the digital signal and a galvanically isolated DC power supply within
the DLS100.
Unlike some other multi-channel isolators, this digital isolator can be
located at the site of the experiment, allowing the use of short
connecting leads and thereby preserving high isolation and fast signal
rise and fall times.
The DLS100 operates in two modes: current source or voltage
source. In the current/voltage source mode, the output current/voltage
is proportional to the amplitude and polarity of the signal generated by
the DS8000. Over-range can occur when the resistance of the load (the
experiment) is too high for the current or voltage that is demanded from
the DLS100.
| DLS100 Specifications |
| Current Source Mode |
| Full-scale* Current |
10 mA, 100 µA, 10 µA, bipolar |
| Compliance Voltage |
± 100 volts |
| Output Impedance |
Greater than 100 Megohms |
| Zero-signal Leakage |
Less than 0.01% of full-scale range setting |
| Linearity |
Better than 0.05% of full-scale range setting |
| Bandwidth |
Range and load dependant: 20 kHz with 10K load and 100 µA or above range.** |
| Voltage Source Mode |
| Full-scale* Voltage |
± 100 volts |
| Max. Current |
1o mA |
| Output Impedance |
Less than 1 ohm |
| Zero-signal leakage |
Less than 1 mV |
| Linearity |
Better than 0.05% of full-scale range setting |
| Bandwidth |
50 kHz |
| Isolation |
| Resistance |
Greater than 1000 Megohms |
| Capacitance |
Less than 10 pF, from output terminals to DS8000 and earth ground |
| Power Requirements |
+ 12 volts and +5 volts, supplied by DS8000 |
| Output Terminals |
Mini-banana jacks |
Activated by conventional logic-level commands, Model A365 can be gated
by any pulse generator, stimulator, or computer output. A tone sounds
when an open electrode circuit is detected or when system compliance is
reached. A second optional tone sounds when a signal is applied to the
input. A test switch is also provided to check battery charge. Stimulus
currents are set using a three-digit control knob and a three-position
range switch. Output current tracks control settings to better than 1%.
Output current is load independent; voltage sufficient to push the
desired current through the load is automatically developed, subject
only to compliance limits. Model A365 produces up to 10 milliampere
current, in three ranges, at more than 100 volts compliance. Output
polarity is determined by a three-position switch on the front panel
(+/-/off). Bipolar current is toggled by the command waveform, setting
alternating pulses as positive or negative. The rechargeable A365R is
supplied with a nickel metal hydride battery stack. The A362 Battery
Charger is required with the A365R. See A365RC, which includes both the A365R and the A362 battery charger.
| OUTPUT WAVEFORM |
DC or current pulse |
| OUTPUT CURRENT RANGES |
0.1, 1.0, and 10mA |
| CURRENT AMPLITUDE ERROR |
0.5% of full scale, max. |
| CURRENT RESOLUTION |
0.1% of full scale, typical |
| OUTPUT LOAD VOLTAGE EXCURSION (COMPLIANCE) |
100V |
| EXTERNAL COMMAND THRESHOLD |
2.2V at 2.6mA, min. 8.5V, max. |
| OUTPUT POLARITY |
Reversible, manual switch or automatic |
| CURRENT RISE TIME & DELAY |
6μs, typical (1KΩ load) |
| CURRENT FALL TIME & DELAY |
10μs, typical (1KΩ load) |
| OUTPUT TO GROUND RESISTANCE |
1012Ω |
| OPTOCOUPLER |
2500V, rated min. breakdown voltage |
| POWER: Model A365D (dry cell) |
16 alkaline 9V batteries, included |
| POWER: Model A365R (rechargeable) |
16 rechargeable NiMH 9V batteries incl. |
| DIMENSIONS |
8.5 x 3.5 x 5 in. (22 x 9 x 12 cm) |
| SHIPPING WEIGHT |
4 lb. (1.8 kg) |
Isostim™Stimulator/Isolator can now come with NiMH batteries in a
rechargeable version. It combines the ease of use and accuracy of WPI's
300 Series stimulators with the power output of a stimulus isolator.
External/DC mode converts Isostim™ to a passive stimulus isolator.
The A320D model is powered by readily obtainable 9-volt alkaline
batteries (included). The rechargeable A320R is supplied with a nickel
cadmium battery stack that provides 10-12 hours of operation before
recharge is required. The A362 battery charger must be used with the
A320R.
WPI also offers high current stimulus isolator, which combines optical
isolation with a ±100 mA current generator. A365 model delivers
positive, negative, or bipolar currents. The input connector is a
standard BNC, allowing signals from any source - such as computer D/A or
I/O lines - to be used.
A385 rechargeable version is not appropriate for transcutaneous
stimulation, and the A382 system charger must recharge the batteries.
Indicator lights and audible alarms keep the user constantly aware of
battery charge status.
WPI's linear stimulus isolator model A395, also available in
rechargeable version will replicate a programmed waveform of any shape
or polarity. Battery operated, and photoelectrically-isolated from the
input voltage drive, the instrument regenerates as output currents the
original waveforms provided by your D/A converter or signal generator.
| OUTPUT WAVEFORM |
DC or current pulse |
| OUTPUT CURRENT RANGES |
0.1, 1.0, and 10mA |
| CURRENT AMPLITUDE ERROR |
0.5% of full scale, max. |
| CURRENT RESOLUTION |
0.1% of full scale, typical |
| OUTPUT LOAD VOLTAGE EXCURSION (COMPLIANCE) |
100V |
| EXTERNAL COMMAND THRESHOLD |
2.2V at 2.6mA, min. 8.5V, max. |
| OUTPUT POLARITY |
Reversible, manual switch or automatic |
| CURRENT RISE TIME & DELAY |
6μs, typical (1KΩ load) |
| CURRENT FALL TIME & DELAY |
10μs, typical (1KΩ load) |
| OUTPUT TO GROUND RESISTANCE |
1012Ω |
| OPTOCOUPLER |
2500V, rated min. breakdown voltage |
| POWER: Model A365D (dry cell) |
16 alkaline 9V batteries, included |
| POWER: Model A365R (rechargeable) |
16 rechargeable NiMH 9V batteries incl. |
| DIMENSIONS |
8.5 x 3.5 x 5 in. (22 x 9 x 12 cm) |
| SHIPPING WEIGHT |
4 lb. (1.8 kg) |
|