Products Menu

like us on facebook New stuff, articles, promos

Tech Support and Returns

DNA/RNA Quantification Using SPT2 and a Tidas Spectrometer Print E-mail
Written by Alexander Dickson   

AbstractSPT-2

Concentrations of DNA in solution (31µg/mL and 144µg/mL) were measured with a spectrometer and UV/VIS light source in a SPT-2. Due to the 1cm pathlength, use of a SPT-2 does not require a pre-measurement dilution within this concentration range, thus a potential source of error was eliminated. Due to the design of the SPT-2, routine measurements can be taken with samples of 10µL in standard 200µL PCR tubes. If caution is taken with regard to the location of the probe tip within a PCR via, measurements remain repeatable in sample sizes as small as 5µL.
Read more....

Experimental Procedure

Standard solutions of DNA (Sigma D1626) were prepared gravimetrically using 18.2MΩ/cm ultrapurified water as a solvent. Solutions were prepared between 0.0µg/mL and 143.9µg/mL.

Measurements were taken in triplicate using an SPT-2. The SPT-2 was connected to a Tidas I spectrometer (WPI #TIDAS-1) and a UV/VIS light source (WPI #D4H).

Data were collected in 1nm increments across the full range of the instrument (190nm-720nm). The instrument was configured such that reference measurements yielded an 80% total intensity. All measurements utilized 18.2MΩ/cm ultrapurified water as a reference solution.

Results

Experimental results are presented in Table 1:

DNA [µg/mL]
Absorbance @260nm  [AU]
0.00
-0.0061 ± 0.0001
30.83 0.5745 ± 0.0011
56.59 1.0309 ± 0.0020
85.49 1.4497 ± 0.0016
115.66 1.7170± 0.0025
143.94 1.8280 ± 0.0006

Table 1: DNA Concentrations and Resultant Absorbance Values

Since absorbance with respect to concentration follows the Beer-Lambert Law,
A= εlc
expected absorbance values were calculated from the DNA solution concentrations. Literature values of ε for dsDNA are listed as 0.020µg/ml*cm.

Experimental data can be found in Figure 1 with the calculated absorbance measurements indicated by a solid line. Absorbance measurements are expressed at a 260nm wavelength. Deviation from the theoretical value at higher absorbance values is a result of stray light interference within the spectrometer.

Measured v. theoretical absorbance

Figure 1: Measured versus Theoretical Absorbance

Typical DNA measurment  

Figure 2: Typical DNA Measurement (56.6µg/mL)

PDF version of this document

For more information on WPI spectroscopy equipment, call 866.606.1974 (US only) or email This e-mail address is being protected from spam bots, you need JavaScript enabled to view it .

Last Updated ( Monday, 16 May 2011 )
 
< Prev   Next >