Abstract
Concentrations of DNA in solution (31µg/mL and 561µg/mL) were measured with a spectrometer and UV/Vis light source in a cuvette. A 2mm pathlength cuvette does not require a pre-measurement dilution within this concentration range, thus a potential source of error was eliminated. Although a 2mm cuvette has a total internal volume of 0.7mL, only 350µL is required to obtain an accurate measurement.
Read more...
Experimental Procedure
Standard solutions of DNA (Sigma D1626) were prepared gravimetrically
using 18.2MΩ/cm ultrapurified water as a solvent. Solutions were
prepared between 0.0µg/mL and 561.1µg/mL.
Measurements were taken in triplicate using a 2mm cuvette (WPI #CUV2102-1) in a standard cuvette holder (WPI #89340). An appropriate cuvette spacer (WPI #89342)
allowed for consistent placement of the cuvette within the holder. The
cuvette holder was attached via two 500µm SMA-terminated fiber optic
cables (WPI #FO-500SMA1M) to a Tidas I spectrometer (WPI #TIDAS-I) and a UV/Vis light source (WPI #D4H).
Data were collected in 1nm increments across the full range of the
instrument (190nm-720nm). The instrument was configured such that
reference measurements yielded an 80% total intensity. All measurements
utilized 18.2MΩ/cm ultrapurified water as a reference solution.
Results
Experimental results are presented in Table 1:
DNA [µg/mL]
|
Absorbance @260nm [AU]
|
| 0.00 |
0.002 ± 0.001 |
| 30.83 |
0.101 ± 0.009 |
| 56.59 |
0.200 ± 0.004 |
| 85.49 |
0.311 ± 0.005 |
| 115.66 |
0.409 ± 0.003 |
| 143.94 |
0.497 ± 0.009 |
| 281.63 |
0.879 ± 0.010 |
| 428.91 |
1.243 ± 0.012 |
561.15
|
1.555 ± 0.003 |
Table 1: DNA Concentrations and Resultant Absorbance Values
Since absorbance with respect to concentration follows the Beer-Lambert Law,
A= εlc
expected absorbance values were calculated from the DNA solution
concentrations. Literature values of ε for dsDNA are listed as
0.020µg/ml*cm.
Experimental data can be found in Figure 1 with the calculated
absorbance measurements indicated by a solid line. Absorbance
measurements are expressed at a 260nm wavelength. Deviation from the
theoretical value at higher absorbance values is a result of stray light
interference within the spectrometer.
Figure 1: Measured versus theoretical abosrbance.
Figure 2: Typical DNA measurement (281.6µg/mL).
PDF version of this document